AccuPulse NIBP Simulator Technical Specifications

AccuPulse
AccuPulse NIBP CalTables™
Datascope Accutorr Plus
GE Medical Dash 4000
GE Medical Dinamap Pro
Philips Medical M3
Welch-Allyn Propaq Encore
Philips Medical VS1
Colin Medical
Criticare
CAS Medical
SunTech
Please contact the factory for the availability of other calibration tables.
BP Presets
Systolic/Diastolic (MAP):
ADULT:
250/190 (212)
200/150 (169)
150/100 (119)
120/80 (95)
100/70 (82)
80/50 (62)
60/30 (42)
NEONATAL:
150/120 (129)
120/90 (99)
100/70 (79)
80/50 (59)
60/30 (39)
35/15 (21)
Exact Diastolic, Systolic and MAP values are a function of the specific CalTables™.The BP Presets above are taken from the Welch-Allyn Propaq Encore CalTable™.
Accuracy:
<± 0.5% of selected target value
Oscillometric Pulse  
Pulse Amplitude @ MAP: 2.0 mmHg
Pulse Volume Range: 0 to 4.4 ml
Pulse Rise Time: 80 msec (minimum)
   
Pulse Rate  
Pulse Rate Range: 15-330 bpm
Pulse Rate Accuracy: ±1 bpm, 15-200 bpm,
±2 bpm 201-330 bpm
   
Displayed/MeasuredParameters
Dynamic NIBP Cuff PressureWaveform
Measurement Time: 0.0 to 180.0 seconds
Maximum Inflation Pressure: 0 to400 mmHg*
Inflation Time: 0.0 to 180.0 sec
   
Inflation Rate:
0 to 999.9 mmHg/sec
Minimum Pressure: 0 to 400 mmHg*
Deflation Time: 0.0 to 180.0 sec
Deflation Rate: -0.0 to -999.9 mmHg/sec
   
Digital Manometer
 
Pressure Range: 0.0 to 400.0 mmHg*
Accuracy: ± 0.5 mmHg, traceable to NIST
Resolution: ± 0.1 mmHg
   
Leak Test  
Start Pressure:
200 mmHg nominal,
10 to 370 mmHg via Remote Control
Pressure Drop: 0 to 400 mmHg*
Elapsed Time: 60 Seconds
High Resolution Manometer: 0.0 to 400.0mmHg*
   
Over-Pressure
Automatically tests operation of the monitor's relief valve
Pop-Off Pressure: 0 to 400 mmHg*
Instaneous Pressure: 0 to 400 mmHg*
   
Autosequences

User can utilize autosequences to test NIBP monitors with
a specific series of AccuPulse
NIBP performance tests.

 
Cuff Supports
Adult and Neonate mandrels are built into AccuPulse for portability and ease of use.
• Plastic and Aluminum Cylinders
• Adult (3.5" OD, 7.25" width)
• Neonate (1.25" OD, 2.25" width)
 
Adapter Hoses

Adapter hoses insert between the NIBP device,
cuff and analyzer. These adapters are compatible
with oscillometric NIBP monitors.
• Male/Female Luer
• Male/Female Clippard (GE Medical, Draeger/Siemens)
• Colder/CPC (GE Medical, Protocol Systems)
• OBAC Quick Release (Philips Medical)
• Universal 5/32" I.D. Hose

 
Self Test Accessories
• Pressure Bulb Assembly (tees into any Cuff Adapter)
• Self system leak test hose (plugged at distal end)
 
Communication Ports
2 RS-232 (second port available Q2 2006)
1 USB Port (available Q2 2006)
 
Display
Non-Glare Graphic LCD
CCFL Backlight
Resolution: 240x320 (1/4 VGA)
Thumbwheel contrast control
 
Power
100-240 VAC, 50 watts, 50-60 Hz, Desktop Switcher
Output: 24VDC @ 2.1A, 6 foot cable
Safety Agency Approvals: UL, CE,TUV
Internal Battery: nickel metal hydride (NiMH)
Voltage: 19.2V
Capacity: 2.5A-Hr, 150 NIBP Simulations
Charge Time: 16 Hours
 
Weight
8.0 pounds with internal battery
 
Dimensions
8" Wide x 5" High x 12" Long
 
Standard Accessories
• External Power Supply and Cord
• Adult and Neonate Cuff Supports (built-in)
• Operation Manual
• Internal Pump
• Internal Battery
 
Optional Accessories
• Carry Case
• Five Adapter Hoses
• Inflation Bulb with “T” fitting
• Self System Leak Test Hose

  *400.0 mmHg typical

 