Title: Cables Required for Testing Datex•Ohmeda Pulse Oximeters
Affected Product(s): SmartSat

Datex-Ohmeda pulse oximeters require different SmartSatSimulator Cables depending on the model to be tested. For the SmartSat “Oximeter” setting, all Ohmeda-Type * pulse oximeters are tested using the Ohmeda mode on the SmartSat's front panel, and all Datex-Type ** pulse oximeters are tested using the Datex mode on the SmartSat's front panel.
No new Probe Test Cables are required for any of these models.

* Note: Ohmeda-Type pulse oximeters are identified by the pulse oximeter probe used. Ohmeda-Type probes use a circular, hard-plastic, 9-pin connector (typically, only 7 pins are populated) with a “press-to-release” locking clip.

** Note: Datex-Type pulse oximeters are identified by the pulse oximeter probe used.
Datex-Type probes use a circular, 10-pin connector (typically, only 7 pins are populated). The Datex-Type circular connector has a larger diameter than the Ohmeda-Type.

Pulse Oximeter Model

SmartSat Simulator Cable
Datex-Ohmeda TuffSat “TUFFSAT”
Datex-Ohmeda 3800, 3900 “DATEX-OHMEDA”
Datex-Ohmeda 3775 “DATEX-OHMEDA” or “OHMEDA”
Datex-Ohmeda S/5 Anesthesia Monitor “DATEX”
Datex-Ohmeda Cardiocap “DATEX”
Datex-Ohmeda Capnomac “DATEX”
Ohmeda 3700 (Legacy Model) “DATEX-OHMEDA” or “OHMEDA”
Datex SATLITEtrans (Legacy Model) “DATEX”
Ohmeda TuffSat (no PI r on the front panel)
(Original/Legacy TuffSat w/o perfusion index)
“M-3000” †

† We do not stock this adapter, as Datex-Ohmeda states that limited numbers of the
Original TuffSat are in the field.; contact Clinical Dynamics if you believe you require it.

Contact Clinical Dynamics to obtain these SmartSat Simulator Cables.

If you have any questions, please contact Clinical Dynamics: service@clinicaldynamics.com

U.S. Customers Please Call Toll Free: 800.247.6427
International Customers Please Call: 203.269.0090
Fax: 203.269.3402

Clinical Dynamics Corporation
10 Capital Drive
Wallingford, CT 06492


Clinical Dynamics Corporation, 10 Capital Drive, Wallingford, CT 06492 U.S.A.
In U.S.A.: 800.247.6427 International: 203.269.0090 Fax: 203.269.3402 E-Mail: info@clinicaldynamics.com
©2005 Clinical Dynamics Corporation