Datex-Ohmeda pulse oximeters require different SmartSat™ Simulator Cables depending on the model to be tested. For the SmartSat “Oximeter” setting, all Ohmeda-Type * pulse oximeters are tested using the Ohmeda mode on the SmartSat's front panel, and all Datex-Type ** pulse oximeters are tested using the Datex mode on the SmartSat's front panel.
No new Probe Test Cables are required for any of these models.
* Note: Ohmeda-Type pulse oximeters are identified by the pulse oximeter probe used. Ohmeda-Type probes use a circular, hard-plastic, 9-pin connector (typically, only 7 pins are populated) with a “press-to-release” locking clip.
** Note: Datex-Type pulse oximeters are identified by the pulse oximeter probe used.
Datex-Type probes use a circular, 10-pin connector (typically, only 7 pins are populated). The Datex-Type circular connector has a larger diameter than the Ohmeda-Type.
Pulse Oximeter Model
| SmartSat Simulator Cable |
| Datex-Ohmeda TuffSat | “TUFFSAT” |
| Datex-Ohmeda 3800, 3900 | “DATEX-OHMEDA” |
| Datex-Ohmeda 3775 | “DATEX-OHMEDA” or “OHMEDA” |
| Datex-Ohmeda S/5 Anesthesia Monitor | “DATEX” |
| Datex-Ohmeda Cardiocap | “DATEX” |
| Datex-Ohmeda Capnomac | “DATEX” |
| Ohmeda 3700 (Legacy Model) | “DATEX-OHMEDA” or “OHMEDA” |
| Datex SATLITEtrans (Legacy Model) | “DATEX” |
Ohmeda TuffSat (no PI r on the front panel) (Original/Legacy TuffSat w/o perfusion index) | “M-3000” † |
† We do not stock this adapter, as Datex-Ohmeda states that limited numbers of the
Original TuffSat are in the field.; contact Clinical Dynamics if you believe you require it.
Contact Clinical Dynamics to obtain these SmartSat Simulator Cables.
If you have any questions, please contact Clinical Dynamics: service@clinicaldynamics.com